Agilent Technologies E8285A manual Output 714MEASAFRDISTNHLIMSTAT?

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 41
Image 41

Chapter 1, Description of Keys

Keys

To Turn High and Low Measurement Limit Checking ON and OFF. Use the :HLIMit:STATe and :LLIMit:STATe commands to turn high and low measurement limit checking ON and OFF.

Syntax

:HLIMit:STATe ON

:HLIMit:STATe OFF

:LLIMit:STATe ON

:LLIMit:STATe OFF

Example

OUTPUT 714;"MEAS:AFR:DISTN:HLIM:STAT ON"

OUTPUT 714;"MEAS:AFR:DISTN:LLIM:STAT ON"

This turns high and low measurement limit checking ON for the AF Analyzer dis- tortion measurement.

To Query the State of High and Low Measurement Limit Checking. Use the :HLIMit:STATe? and :LLIMit:STATe? commands to query the current state of the high and low measurement limit checking. The returned value is either: 0 (OFF) or 1 (ON).

Syntax

:HLIMit:STATe?

:LLIMit:STATe?

Example

OUTPUT 714;"MEAS:AFR:DISTN:HLIM:STAT?"

ENTER 714;Hi_state ! 1 = ON, 0 = OFF

OUTPUT 714;"MEAS:AFR:DISTN:LLIM:STAT?"

ENTER 714;Lo_state ! 1 = ON, 0 = OFF

This queries the state of high and low measurement limit checking for the AF

Analyzer distortion measurement.

To Set High and Low Measurement Limits. Use the :HLIMit:VALue and :LLIMit:VALue commands to set the high and low measurement limit values.

Syntax

:HLIMit:VALue

:LLIMit:VALue

Example

OUTPUT 714;"MEAS:AFR:FM:HLIM 7.5 KHZ"

OUTPUT 714;"MEAS:AFR:FM:LLIM 2.5 KHZ"

This sets a high measurement limit of 7.5 kHz and a low measurement limit of 2.5 kHz for the AF Analyzer FM deviation measurement.

NOTE:When setting high and low limit values, a non–Attribute Unit unit-of-measure must be specified in the command string, otherwise the current Attribute Unit is assumed by the Test Set. Refer to “To Specify Units-of-Measure for GPIB Data Transfer” in the Operating Overview chapter of the E8285A User’s Guide .

41

Page 41
Image 41
Agilent Technologies E8285A manual Output 714MEASAFRDISTNHLIMSTAT?