Chapter 1, Description of Keys

Keys

To Query the State of the Meter.

Use the :METer:STATe? commands to query the state of the analog bar-graph meter. The query returns a 1 if the meter is ON, and a 0 if the meter is OFF.

Syntax

:METer:STATe?

Example

OUTPUT 714;"MEAS:RFR:POW:MET:STAT?"

ENTER 714;Meter_on_off ! returns a 1 (ON) or 0 (OFF)

This queries the state of the analog bar-graph meter for the TX power measurement.

To Set the Number of Intervals on the Meter.

Use the :METer:INTerval commands to set the number of intervals displayed on the analog bar-graph meter.

Syntax

:METer:INTerval <integer valve>

Example

OUTPUT 714;"MEAS:RFR:POW:MET:INT 5"

This sets the number of intervals displayed on the analog bar-graph meter for the TX power measurement.

To Query the Number of Intervals on the Meter.

Use the :METer:INTerval? commands to query the number of intervals displayed on the analog bar-graph meter.

Syntax

:METer:INTerval?

Example

OUTPUT 714;"MEAS:RFR:POW:MET:INT?"

ENTER 714;Num_intervals

This queries the number of intervals displayed on the analog bar-graph meter for the TX power measurement.