Chapter 4, Description of Fields

Fields

Center Freq (CDMA)

This field sets the center frequency of the CDMA spectrum analyzer. This field is displayed when the Controls field is set to Main and the RF Display

field on the CONFIGURE screen is set to Freq.

GPIB Example

“RFAN:FREQ 850MHZ“

sets the center frequency to 850 MHz.

See Also

RF Display field description, on page 367

Screens Where Field is Present

CDMA REVERSE CHANNEL SPECTRUM

Change

This field defines the type of power control bits that will be sent to the mobile station when Execute is selected. The range over which the mobile station will be commanded to change its power level is determined in the Steps field.

n Up sends a series of “0” power-control bits, causing the MSUT to increase its power, then returns to the "normal" sequence.

sends a series of “1" power-control bits, causing the MSUT to decrease its power, then returns to the "normal" sequence.

Ramp sends a series of "n up" followed by a series of "n down" bits.

GPIB Example

“DISP CTXR;CDMA:PCONtrol:CHANge ‘n up’”

selects “n up” power control steps.

Operating Considerations

See "Closed Loop Pwr Cntl" on page 203.

Screens Where Field is Present

CDMA TRANSMITTER POWER RANGE TEST