Chapter 4, Description of Fields

Fields

Open Loop Time Response (Execute)

This field is displayed on the OPEN LOOP TIME RESPONSE screen when “Main” is selected in the "Controls (Open Loop Time Response)" field on page 212. Selecting this field executes a time response to open loop power control test.

GPIB Example

“MEAS:CDManalyzer:OLTR:EXECute”

executes a time response to open loop power control test.

Operating Considerations

When Execute is selected, the Test Set will step its output power up or down depending on the value entered in the "Step Size" field on page 414. The mobile station’s output power as a function of time will be graphically displayed on the OPEN LOOP TIME RESPONSE screen.

Screens Where Field is Present

OPEN LOOP TIME RESPONSE

Options

This field is used to enter the instrument’s option number or numbers if any.

Operating Considerations

This field may be left blank, or otherwise may include other calling name options, for example:

Printer options – LN=#,START,END

Where # is the number of lines on each page.

Where START causes a form feed at the start of each printout. Where END causes a form feed at the end of each printout.

Data collection options – NN

Where NN is the number of records (file size) for the mass-storage location where data will be collected on disk or memory card. The default record size is 80.

Screens Where Field is Present

TESTS (External Devices)

320

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 320