Chapter 4, Description of Fields

Fields

Then, to measure maximum power, the Test Set sets Sector A power to -104 dBm/ BW, turns the Call Drop Timer “Off”, and sends “Always Up” power control bits to the mobile station, using a procedure similar to the EIA/TIA IS-98 test “Maximum RF Output Power”. An Average Power measurement is made, and the Test Set is returned to closed loop power control mode.

See Also

Max Power field description, on page 293

Min Power field description, on page 296

Screens Where Field is Present

CDMA TRANSMITTER POWER RANGE TEST

Mobile Parm

These fields allow entry or display of a mobile station parameter identifier or value.

MUX1_REV_1 through MUX1_REV_14

MUX2_REV_1 through MUX2_REV_25 (TSB-74 and J-STD-008 only)

MUX1_FOR_1 through MUX1_FOR_ 14

MUX2_FOR_1 through MUX2_FOR_26 (TSB-74 and J-STD-008 only)

PAG_1 through PAG_7

ACC_1 through ACC_8

LAYER2_RTC1 through LAYER2_RTC5

SYS_TIME

GPIB Example

“DISP CFER;CDMA:MOBile:PARM ‘MUX1_REV_1’“

identifies the mobile station parameter MUX1_REV_1.

“DISP CFER;CDMA:MOBile:PARM:VALue 0“

sets the mobile station’s MUX1_REV_1 parameter value to 0.

Operating Considerations

This field allows selection of a mobile station Parameter Identifier, and displays the value of the selected parameter. If a call is connected and a value is entered, an attempt will be made to write this value into the corresponding register in the mobile station. This field is only valid for mobile stations that conform to the EIA/ TIA IS-95-A standard.

Screens Where Field is Present

CDMA MOBILE REPORTED FER

298

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 298
Image 298