Chapter 4, Description of Fields

Fields

Output Port

This field selects the RF output port on the Test Set.

RF Out selects the RF IN/OUT port on the Test Set’s front panel and displays the RF Out field for entering CDMA forward channel path loss.

Dupl selects the DUPLEX OUT port on the Test Set’s front panel and displays the Duplex Out field for entering CDMA forward channel path loss.

GPIB Example

“CONF:OUTP ‘Rf Out’”

selects the RF Output port.

Operating Considerations

Maximum signal levels at each port are printed on the front panel. The Duplex

Out port provides higher power.

CAUTION:Applying reverse RF power to the DUPLEX OUT connector can damage the instrument. (A message is displayed when an overpower conditions occurs.) Whenever possible, use the RF IN/ OUT connector when testing transceivers to prevent damage from accidental transmitter keying.

If a reverse-power condition triggers the internal protection circuit, remove the reverse-power signal and press the Meas reset key or turn the Test Set off and on to reset it.

Screens Where Field is Present

CONFIGURE

CDMA REVERSE CHANNEL SPECTRUM

SPECTRUM ANALYZER

Output Results For

This field selects which test results are output. The underlined entry is the active choice.

All outputs all test results on to the location selected in the Output Results To: field.

Failures outputs only the results of tests which fail.

Screens Where Field is Present

TESTS (Execution Conditions, Printer Setup)

324

S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp

Page 324
Image 324
Agilent Technologies E8285A manual Output Port, Output Results For, This field selects the RF output port on the Test Set