Agilent Technologies E8285A manual Programming Meas reset

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 44
Image 44

Chapter 1, Description of Keys

Keys

To Reset Measurement Limit Detection. Use the :HLIMit:RESet and :LLIMit:RESet commands to reset measurement limit detection. Once a high or low measurement limit has been exceeded (:HLIMit:EXCeeded? returns a 1 or :LLIMit:EXCeeded? returns a 1), measurement limit detection is disabled until reset by the :RESet command.

Syntax

:HLIMit:RESet

:LLIMit:RESet

Example

OUTPUT 714;"MEAS:AFR:FM:HLIM:RES"

OUTPUT 714;"MEAS:AFR:FM:LLIM:RES"

This resets high and low measurement limit detection for the AF Analyzer FM de- viation measurement.

See Also

“Setting Measurement Limits” in the Operating Overview chapter of the E8285A User’s Guide .

Meas reset

Meas reset clears the measurement history for all of the instrument’s measurement algorithms (such as the averaging and peak hold functions) to restart all measurements that are in progress.

Programming Meas reset

The GPIB commands :MEASure:RESet are used to select this function programmatically.

Example

OUTPUT 714;":MEAS:RES"

This resets all of the active measurements in the Test Set.

44

S:\agilent\e8285\REFGUIDE\MANUAL\keys.chp

Page 44
Image 44
Agilent Technologies E8285A manual This resets all of the active measurements in the Test Set, Programming Meas reset