Agilent Technologies E8285A manual Programming Meter, To Turn the Meter on and OFF

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 45
Image 45

Chapter 1, Description of Keys

Keys

Meter

The Meter function displays measurements graphically. The Meter format is available for most measurements. To determine if the Meter format is provided for a measurement, position the cursor in front of the measurement’s units field and press the knob. If the message “Press ON/OFF, LIMITs, REF, AVG, METER, or units” is displayed, the meter format is provided.

Programming Meter

The GPIB command :METer is used to select this data function programmatically.

To Turn the Meter ON and OFF.

Use the :METer:STATe commands to turn the meter ON and OFF. The parameter can be a 1 or ON to turn the meter on and a 0 or OFF to turn the meter off.

Syntax

:METer:STATe <ON> or <1> :METer:STATe <OFF> or <0>

Example

OUTPUT 714;"MEAS:RFR:POW:MET ON"

This turns the analog bar-graph meter ON for the TX power measurement.

45

Page 45
Image 45
Agilent Technologies E8285A manual Programming Meter, To Turn the Meter on and OFF