Chapter 2, Description of Connectors

Connectors

10 MHz OUTPUT

The 10 MHz output is one output of a reference phase-locked-loop that generates 10 MHz reference signals for all of the digital and analog phase-locked-loops in the Test Set. This reference phase-locked-loop can lock to many frequencies, which are listed in the description of the EXT REF INPUT connector.

Operating Considerations

BNC

This reference signal will be phase-locked to the reference input if the operating conditions specified in the EXT REF INPUT connector description are met.

Nominal output level is >+5 dBm

Output impedance is 50 Ω

Frequency stability is the same as the signal applied to the EXT REF INPUT connector.

See Also

"EXT REF INPUT" on page 73

80

S:\agilent\e8285\REFGUIDE\MANUAL\connectr.chp

Page 80
Image 80
Agilent Technologies E8285A manual MHz Output, See Also EXT REF Input on