Agilent Technologies E8285A manual Display Interim Results, Display User Mssgs

Models: E8285A

1 464
Download 464 pages 57.8 Kb
Page 231
Image 231

Chapter 4, Description of Fields

Fields

Display Interim Results

This field allows the display of the current FER, errors, and number of frames counted while a measurement is in progress.

Yes displays periodic updates from the cell site’s processor.

No display one measurement result at the end of the test.

GPIB Example

“DISP CRXT;DISPlay:FERasure:INTerim:RESults ’YES’“

displays interim FER results.

Operating Considerations

This field must have Yes selected to display interim results.

If Display Interim Results is Yes, and a measurement result is queried over GPIB during a measurement, an interim measurement will be returned.

Screens Where Field is Present

CDMA CELLULAR MOBILE RECEIVER TEST

Display User Mssgs

This field allows the user to turn off display of messages that appear on the front panel display when Test Set the current FER, errors, and number of frames counted while a measurement is in progress.

Yes enables messages to be displayed.

No disables user message display.

GPIB Example

“DISP CONF;DISPlay:MESSages ’YES’“

enables displays of user messages.

Screens Where Field is Present

CONFIGURE

231

Page 231
Image 231