File – Import – Digitizer –Step Scan (When VIVID 700 is Selected)

9 Place the calibration chart on the rotating stage.

Position the chart with the black line facing the instrument

and place it on the rotating stage so that both entire side surfaces are observed in the view- finder.

If “Chart” is not displayed in [Current Angle], click the left mouse button and select “Chart” from the pull-down menu that appears.

Chapter 2

File

Menu

10 Click the [OK] button.

The calibration chart will be scanned.

Note

Make sure that the acquired range image looks like the good example given below. If it resembles the bad example, click the [Scan] button again to re-scan the calibration chart.

Memo

The calibration chart can be scanned at any angle of the rotating stage.

Good example

Bad example

 

 

 

 

11 Click the [Convert] button.

The [File-Import-Digitizer-Step Scan-Convert] dialog box will appear.

700

108