File – Import – Digitizer – Step Scan (When VIVID 900/910 is Selected)

Chapter 2

File

Menu

nWhen Using the Bench Top Frame Set

Memo

The VIVID 910 or VIVID 900 can be mounted on the bench top frame set vertically or horizontally. For the mounting method, refer to the instruction manual of the frame set.

Saving the calibration chart data helps you reduce work time when scanning different objects under the same conditions.

Operating Procedure

1

From the [File] menu, select [Import],

• The monochrome monitor image currently cap-

 

[Digitizer] and then [Step Scan].

tured by the VIVID 910 or VIVID 900 will ap-

 

The [File-Import-Digitizer-Step Scan] dialog

pear in the image area of the dialog box.

 

box will appear.

 

2 Select the desired installation direction of the VIVID digitizer.

From the [Hardware] tab, check the [Bench Top Frame set] checkbox, and select the VIVID digitizer installation direction by clicking the corresponding [Mounted] radio button.

Select [Vertical] if the VIVID 910 or VIVID 900 is mounted vertically, or select [Horizontal] if it is mounted horizontally.

Memo

If the VIVID digitizer is mounted horizontally, the preview image will be in portrait form.

After the image is converted to 3D, it will be displayed according to the coordinate system of the VIVID.

910

900

3 PlaceNote the object on the rotating stage.

If the VIVID digitizer is mounted on the frame set hori- zontally, the object will be affected by repeated reflection on the rotating stage surface, preventing correct scan operation. To prevent this, place an appropriate item of the following thickness under the object to raise it.

Measurement Distance

Thickness of Item

 

 

600 mm

Approx. 30 mm

 

 

1000 mm

Approx. 20 mm

 

 

4 To display the object in the middle of the image area, change the position of the object or move the instrument back and forth to change the view angle.

Memo

If necessary, replace the lens attached to the VIVID digitizer.