File – Import – Digitizer – Step Scan (When VIVID 900/910 is Selected)
Chapter 2
File
Menu
10 Place the calibration chart on the rotating stage.
When the “Scanning the Calibration Chart.” message dialog box appears, position the chart with the black line facing the VIVID, and place it on the rotating stage so that both entire side surfaces are observed in the view- finder.
•If “Chart” is not displayed in [Current angle], click the left mouse button and select “Chart” from the
Calibration chart S
Memo
If it is not possible to perform calibration using calibration chart S, use calibration chart M.
#ALIBRATION #HART -
910
900
11 Click the [OK] button. |
|
| |
The calibration chart will be scanned. | Good example | Bad example | |
Note | |||
|
| ||
|
|
Make sure that the acquired range image looks like the good example given below. If it resembles the bad example, click the [Scan] button again to
12 Click the [Convert] button.
The