File – Import – Digitizer – One Scan (When VIVID 9i is Selected)

11 Click the [Scan] button.

At this time, the data in the Work window area

does not correspond to the one in the Store RED window area, so a red frame is displayed.

Chapter 2

File

Menu

12 Designate three or more corresponding points. Set the corresponding points in the Work window area and then store window alternately. When the corresponding points have been designated properly, the red frame will turn to blue.

If desired points are set by left-clicking them, registration will be performed each time 3 or more pairs of corresponding points are set.

By using middle button or left button + [Shift] button, registration can be performed when 1 or more pairs of corresponding points are set.

Memo

Registration with 1 pair of corresponding points must be used for objects with characteristic shape.

Since registration may not be performed successfully for objects with non-characteristic shape (e.g. flat planes, cylindrical surfaces), using 3 pairs of corresponding points is recommended

Each time corresponding points are designated, registration among data is performed.

If the [Auto Store] checkbox has been checked, proceed to Step 13 after the corresponding points are designated properly.

Left

: Sets a point.

 

 

(Registration is performed using 3 or more pairs of corresponding points)

Middle

: Sets a point.

([Shift] + Left)

(Registration is performed using 1 or more pairs of corresponding points)

Right

: Points in a Work window area or Store window area can be de-

 

 

leted by clicking them in the reverse order in which they were set.

 

 

Blue

 

 

 

 

 

 

 

 

 

9i

1

 

 

2

 

 

 

 

 

3

4

 

 

 

 

5

 

6

 

 

 

 

37