File – Import – Digitizer – Step Scan (When VIVID 900/910 is Selected)

11 Place the calibration chart on the rotating stage.

Position the chart with the black line facing the VIVID digitizer, and place it on the rotating stage so that both entire side surfaces are observed in the viewfinder.

If “Chart” is not displayed in [Current angle], click the left mouse button and select “Chart” from the pull-down menu that appears.

12 Click the [OK] button.

The calibration chart will be scanned, and the “Save the chart data ?” message dialog box will appear.

To save the calibration chart data, click the [OK] button.

Note

If accurate data is required, it is recommended that calibration be performed each time.

Make sure that the acquired range image looks like the good example given below. If it resembles the bad example, click the [Scan] button again to re-scan the calibration chart.

Calibration chart S

Memo

If it is not possible to perform calibration using calibration chart S, use calibration chart M.

Chapter 2

#ALIBRATION #HART -

File

Menu

 

 

 

 

 

 

Memo

 

 

910

Saving the calibration chart data helps you reduce work

900

time when scanning different objects under the same

 

conditions.

 

Good example

Bad example

 

 

 

 