Click the [Calibration] button.
The program executes calibration. If calibration is successful, the “Calibration success.” message dialog box will appear.
Click the [OK] button.
The dialog closes. This completes the calibration.
File – Import – Digitizer – Easy Align
•If calibration is not successful, the “Calibration failure.” massage dialog box appears. If this hap-
pens, click the dialog’s [OK] button, and effect |
|
the measures listed below in accordance with the | Chapter 2 |
content given in the dialog. | File |
| Menu |
In the case of that the “Lighting is too bright.” |
|
message dialog box will appear, the light source |
|
is too strong, so decrease its intensity. |
|
In the case of that the “Lighting is too dark.” mes- |
|
sage dialog box will appear, the light source is too |
|
dark, so increase its intensity. |
|
In the case of that the “Chart setting is improper. |
|
or Light source is improper.” message dialog box |
|
will appear, either the Color Chart has been im- |
|
properly set or the light source is unstable. Make |
|
sure that the Color Chart directly faces the VIVID |
|
910 and that it fills the screen completely. Also |
|
check to make sure that the fluorescent light is not |
|
flickering. |
|