File – Import – Digitizer – Step Scan (When VIVID 910 is Selected)

8 Place the Calibration Chart on the rotating stage and scan the Calibration Chart.

Position the chart with the black line facing the VIVID, and place it on the rotating stage so that both entire side surfaces are observed in the viewfinder.

This step is not necessary if the Calibration Chart data has already been imported when this dialog box is opened.

To acquire data from the object first then acquire the calibration chart data, perform steps 8 and 9 after completing step 14. In this case, select cur- rent chart angle from [Current Angle] and click [Scan] button, or click [Chart Scan] button to acquire chart data.

The [Chart Scan] button appears when “Free” is selected for [Rotation Step] in [File-Import- Digitizer-Step Scan-Option] dialog box.

Note

After changing the setting orientation at Convert Parameter – Hardware Parameter on the [File-Import- Digitizer-Step Scan-Option] dialog box, the chart data must be acquired again before measuring.

Calibration chart S

 

 

 

 

 

Memo

 

 

If it is not possible to perform calibration using calibra-

 

tion chart S, use calibration chart M.

 

 

 

 

 

 

• Steps 9 and 10 are not necessary when importing

Chapter 2

 

existing the Calibration Chart data (i.e. data that

File

 

has coordinate parameters). Clicking on the [Im-

Menu

 

 

 

port CDM] button will open the file read dialog

 

 

box.

 

 

The Calibration Chart data (chart∗∗∗∗.cdm) that

 

 

has been saved can be imported.

 

 

q Select the file to be imported.

 

 

w Click on the [Open] button.

 

1

2

9 Click the [Scan] button.

The Calibration Chart will be scanned. When the data is recognized as Calibration Chart data successfully:

qThe Calibration Chart data will be stored under the file name of chart∗∗∗∗.cdm (“∗∗∗∗” are the last four digits of the main unit’s serial num- ber) in the Model folder.

wThe data acquired from the subsequent scans will be coordinate-converted around the rota- tional axis.

e The [AutoScan] checkbox will be displayed.

Note

Make sure that the acquired range image looks like the good example given below. If it resembles the bad example, click the [Scan] button again to re-scan the calibration chart.

An error message will be displayed if the data is not recognized as the Calibration Chart data suc- cessfully.

• When “Free” has been selected using the [Rota-

910

tion Step] of the [Stage Parameter], please use the

 

[Chart Scan] button instead of the [Scan] button.

 

Good example

Bad example

 

 

 

 

 

 

 

61