Chapter 3

Error Messages

Error Message (Alphabetically)

Cause

Action

Please select only same format

The [Open] or [Import] button was

Click the [OK] button to close the mes-

files.

clicked even though more than one file

sage dialog box.

 

of different format had been selected.

 

Please select point(s). (An alarm

The command was selected with no

Click the [OK] button to close the mes-

will be heard if the shortcut key is

points selected.

sage dialog box.

used.)

 

 

Please select points within each

The command was selected even though

Click the [OK] button to close the mes-

selected element. (An alarm will

no points had been selected or points

sage dialog box.

be heard if the shortcut key is

had been selected for only some selected

 

used.)

 

element but not all.

 

 

 

Please select polygon mesh ele-

The command was selected even though

Click the [OK] button to close the mes-

ment( s). (An alarm will be heard if

no elements had been selected or ele-

sage dialog box.

the shortcut key is used.)

ments other than polygon mesh elements

 

 

 

 

had been selected.

 

Please select polygon(s). (An

The command was selected even though

Click the [OK] button to close the mes-

alarm will be heard if the shortcut

no polygons had been selected.

sage dialog box.

key is used.)

 

 

Please select three or more

The command was selected even though

Click the [OK] button to close the mes-

points. (An alarm will be heard if

three or more points had not been

sage dialog box.

the shortcut key is used.)

selected.

 

 

 

 

The right mouse button was clicked

 

 

even though more than three pairs of

 

 

corresponding points had not been

 

 

designated.

 

Same element name exists.

The same element name already exists.

Enter a new element name, and click the

Please enter another name. (“File

 

[OK] button to open the file.

name”)

 

 

Same element name exists. Re-

An attempt was made to enter an ele-

To replace the existing element, click the

place?

ment name that already exists.

[Replace] button.

Same file name exists.

An attempt was made to enter a file

To overwrite the file, click the [Over-

 

name that already exists.

write] button.

SCSI read error.

An error occurred with SCSI connection.

Click the [OK] button to close the mes-

 

A read error from the memory card oc-

sage dialog box.

 

curred in VIVID 900/910 (VI-900/ 910).

 

Select more points. We need

Among selected elements, at least one

Click the [OK] button to close the mes-

more than 500 selected points for

element that does not have a certain

sage dialog box.

registration.

number of selected points exists.

 

 

 

Selected file “∗∗∗∗” may be bro-

An attempt was made to read a file that

To continue to read the file, click the

ken. Continue?

had not been saved properly.

[Continue] button. ∗If the [Continue]

 

 

button is clicked, an abnormal stop of the

 

 

software may occur during reading.

Something wrong with VIVID.

A SCSI error other than SCSI connec-

Click the [OK] button to close the mes-

 

tion error, SCSI command end error and

sage dialog box.

 

ready error due to an AF error occurred.

 

The element has no image/The

The command was selected even though

Click the [OK] button to close the mes-

elements have no image. (An

only the polygon mesh elements that

sage dialog box.

alarm will be heard if the shortcut

were defined without copying images or

∗If both polygon mesh elements with

key is used.)

elements other than polygon mesh ele-

images and those without images are

 

 

ments were selected.

selected, this error message will not be

 

 

displayed for those elements that do not

 

 

contain images.

The element has no image. (An

The selected element did not have any

Click the [OK] button to close the mes-

alarm will be heard if the shortcut

images when the command was se-

sage dialog box.

key is used.)

lected.

 

 

 

262