18. ADJUSTMENT

18.2 Shift Programming of Measured Value

This programming is for slight-shifting a measured value. After programming, measured values are shifted by the programmed amount.

1

Example of shift programming

Shift Programming

Measured value ….850.3

Cursor

 

(1)This programming is necessary for each channel.

(2)The cursor shifts to the least significant digit. Execute this

programming with

%

or

/

.

 

 

 

% x 3

Programming …… 850.0

Cursor

2 Programming Flow Chart

<Example> Changing measured value of CH 1 from 850.3 to 850.0

Operation screen

 

DISP

 

2 sec. or

 

Remarks1

 

more

SHIFT

6

A- Z

[Engineering mode selection]

3 sec. or more

Ref.1

 

 

 

    

Cursor Adjustment

[Selecting adjustment]

 

°

°C

ENTRY

 

F

 

 

  

 

 

Cursor

 

°

°

ENTRY

 

 

 

 

 

 

F

C

 

[Selecting shift]

 

 

 

Ref.2

 

   

 

Cursor

Shift

 

 

 

 

 

 

0

1

Remarks2

[Programming channel]

 

CLEAR

CHART

 

 

 

 

 

 

 

 

 

Cursor

Channel

ENTRY

 

 

 

 

 

 

[To measured value display]

 

 

 

   

 

 

 

 

Measured value

Cursor

 

[Programming shift]

 

/

%

 

   

 

 

 

[To channel programming Programmed value

 

screen]

 

ENTRY

 

 

 

 

 

 

 

 

 

 

 

 

 

Cursor

Channel

SHIFT

SET

 

 

 

 

Remarks3

 

 

END

 

 

 

 

 

 

 

Stored

 

Ref.3

 

 

 

 

-107-

Shift value = 850.0-850.3 = -0.3

Before programming

Before starting this programming, wait for at least 30 minutes after turning on the power supply.

By returning to the

Remarks 1 operation screen

Programmed shifting data is canceled if the screen is returned to the operation screen in the procedures before [Storing].

Remarks 2 Skipped channel does not accept any [ENTRY]

When a skipped channel is programmed, ENTRY operation is not accepted.

Remarks 3 Other channels adjustment

Change the connection to other input terminals to be adjusted and repeat the same procedure from the programming channel display (indicated by a dotted line).

Remarks 4 Initialization of shift data (0)

In zero adjustment or span adjustment

screen, when SHIFT and 0 are

CLEAR

pressed simultaneously and then ENTRY is pressed, the correction data of its channel are initialized.

Reference 1 Engineering mode types

: Memory clear

: Adjustment

: Hardware check

Reference 2 Adjustment types

: Measured value adjustment

: Shift programming for measured value

: Printing position adjustment

: Time-axis adjustment for pens

Reference 3 Storing

Store the programmed shifting data into memory.

Page 111
Image 111