4.CONNECTIONS
4.7 Communications terminals
3RS-422A, RS-485 Connections
necessary in the same as | 2 |
| |
1) |
| ||
Personal computer | Line converter | ||
RD 2
SD 3
ER 4
SG 5
DR 6
RS 7
CS 8
Example of 9-in connector
1 RD
2 SD
3 SG
4
RDA 5
RDB 6
SDA 7
SDB 8
RD200/RD2800 hybrid recorders |
| ||
(1) | (2) | (3) |
|
SDA | SDA | SDA |
|
SDB | SDB | SDB |
|
RDA | RDA | RDA |
|
RDB | RDB | RDB | Termination |
SG | SG | SG | resistance |
|
|
| 100 Ω, 1/4W |
Within 15 m |
| Within a total length of 1.2 km (max. 31 recorders) |
| ||
|
|
|
*For the details of the wiring, refer to 2
2) RS-485
Personal computer
RD 2
SD 3
ER 4
SG 5
DR 6
RS 7
CS 8
Example of
Line converter
1 RD
2 SD
3 SG
4
RDA 5
RDB 6
SDA 7
SDB 8
RD200/RD2800 hybrid recorders
(1) | (2) | (3) |
SA | SA | SA |
SB | SB | SB |
SG | SG | SG |
Termination
resistance
100 Ω, 1/4W
Within 15 m Within a total length of 1.2 km (max. 31 recorders)
*For the details of the wiring, refer to 2
-20-