22. SPECIFICATIONS

6 Operation Conditions

Operation conditions: Refer to the following table

Items

Reference

Normal operation

operation

 

 

 

 

 

 

Ambient

21 to 25°C

0 to 31°C

 

 

31 to

temperature

 

 

50°C*

 

 

 

 

Max. ambient

80%RH

80%RH

 

 

80 to

humidity

 

 

50%RH*

 

 

 

 

Min. ambient

20%RH

20%RH

humidity

 

 

 

 

 

Power supply

100VAC ± 1%

90 to 264VAC

voltage

 

 

 

 

 

Power supply

50/60Hz ± 0.5%

50/60Hz ± 2%

frequency

 

 

 

 

 

 

Left/right

0 to 10°

Posture

Forward

 

tilting

 

 

 

 

 

 

 

Backward

0 to 30°

 

tilting

 

 

 

 

 

 

Warming

-up time

30 min. or more

Not necessary

*Decrease from 80%

RH at 31°C, and

decrease to 50% at

40°C

 

 

 

 

 

Transportation condition:

Ambient temperature/humidity… -20 to 60°C, 5 to 90%RH (without condensation)

Vibration…10 to 60Hz, 4.9 m/S2* Impact…392 m/S2*

*When shipping out from a factory

Storage condition:

Ambient temperature/humidity… -20 to 60°C, 5 to 90%RH

(without condensation)

7 General specifications

Rated power voltage: 100 to 240 VAC, 50/60 Hz

Power consumption: Maximum 60V A

Power failure protection: Programmed parameters stored into EEPROM memory. Clock circuit sustained for minimum 8 years by a lithium battery. (8 hours operation per day)

Insulation resistance:

Between primary and protective conductor terminals …20Mor more at 500 V DC

Between secondary and protective conductor terminals …20Mor more at 500 V DC

Between primary and secondary terminals …20Mor more at 500 V DC

Dielectric strength:

Between primary and protective conductor terminals …1 minute at 1500 V AC

Between secondary and protective conductor terminals …1 minute at 500 V AC

Between primary and secondary terminals …1 minute at 2300 V AC

Primary terminals: Power terminals, Alarm output terminals (MOS relay, mechanical relay “a” contact)

Secondary terminals: Measuring input terminals, Remote contacts terminals, Communications interface terminals , Alarm output terminals (Mechanical relay “c” contact),

Case assembly material: Door Frame …ABS resin Front plate…Polycarbonate Enclosure…Steel

Color: Door … Frame: Black (equivalent to Munsell N3.0) Front plate: Transparent

Enclosure…Gray (equivalent to Munsell N7.0) Mounting: Panel mounting

Terminal screws: Power terminals …M4.0 Terminals other than above …M3.5

Weight: RD200 …Approx. 4.0 kg (with fully options) RD2800 …Approx. 9.0 kg (with fully options)

Dimension: RD200144×144×258 (mm) RD2800288×288×220 (mm)

Panel cutoutRD200138×138 (mm) RD2800281×281 (mm)

Clock accuracy: Within ±2 minutes per 30 day

(Under reference operating conditions, Except errors when power supply is turned on or off)

Packing material: Layered cardboard is used as cushion material

8 Safety Regulations

CE: Conforms to EMC and low voltage directives

UL: UL3111-1 (Approval pending)

CSA (C-UL):CSA …C22.2 No.1010 (Approval pending)

IP: Conforms to IEC529 IP54

-128-

Page 132
Image 132
Omega RD200, RD2800 manual General specifications, Safety Regulations