13. REMOTE CONTACTS

13.1 Remote Contacts Function

This explanation is only for the remote contacts (option).

1 Remote Contacts

(1)The following functions are available with the contact signals at remote contacts terminals (EX1 to 4). However, the functions are limited due to four terminals (EX1 to 4) provided. Moreover, some functions are automatically allocated to a certain terminal Nos.

(2)Programming to allocate the functions to terminal Nos. is necessary.

Remote contacts terminals

EX1 EX2 EX3 EX4 COM

 

Functions

Terminals

Remarks

(1)

Printing on/off and selection of three chart

2 terminals (EX1, EX2)

Programming 3 speeds, See Section 13.3. (Note 1)

 

speeds

 

 

 

(2)

Execution of message printing (No.1 to 5)

4 terminals (EX1 to EX4)

Programming message, See Section 11.10. (Note 2)

(3)

Execution of message printing (No.1 and 2)

2 terminals (EX1, EX2)

Programming message, See Section 11.10. (Note 3)

 

 

 

 

(4)

Execution of digital data printing

Any 1 terminal

 

(5)

Execution of list printing (list 1, 2, 3)

Any 3 terminals

 

 

 

 

 

(6)

Execution of operation record (No.A to D)

Any 4 terminals

Programming operation record, See Section 13.4.

 

 

 

 

(7)

Reset of totallization

Any 1 terminal

Available in totalization (option)

 

 

 

 

(Note 1) Free terminals left are EX3 and EX4. (Note 2) No free terminals are left.

(Note 3) Free terminals left are EX3 and EX4.

2

Functions and Terminal Contact Signals

ON: Short OFF: Open

Functions

Contact signal at terminal

Program 3 chart speeds first. See Section 13.3.

 

Printing on/off and selection

 

With COM

 

of 3 hart speeds

 

EX1

 

EX2

(1) Printing on/off and selection of

 

 

CS1

OFF

 

OFF

3 chart speeds

Printing: On

 

CS2

ON

 

OFF

 

 

 

CS3

OFF

 

ON

 

Printing: Off

ON

 

ON

(Note) Program “Printing on/off” to on with a key operation in advance.

Program messages first. See Section 11.10.

 

Messag

 

 

With COM

 

 

 

 

 

 

e

EX1

EX2

 

EX3

 

 

EX4*

 

No.1

OFF

OFF

 

OFF

 

For trigger

 

No.2

ON

OFF

 

OFF

 

 

 

 

 

(2) Execution of message printing

No.3

OFF

ON

 

OFF

 

 

 

 

 

 

 

 

 

 

 

No.4

ON

ON

 

OFF

 

 

 

 

 

(No. 1 to 5)

 

 

 

 

 

 

 

 

 

 

 

 

No.5

OFF

OFF

 

ON

 

1 sec. or more

 

 

 

*When the trigger signals is sent (for 1 sec. or more) after selecting the message No., the printing of programmed message starts.

(Note) Program “Printing on/off” to on with a key operation in advance. Execution of message printing with key operation is also available. See Section 9.4.