7. PROGRAMMING

7.3 List of Programming Items

1

Parameters to be Programmed First

Parameters

Key operations

Programming

Defaults

Programming ranges

Ref.

items

page

 

 

 

 

Range/

 

 

 

Range No.

07

01 to 10, 21 to 56, 70 to 80

 

 

+

7

RJ enable/

0

0 (disable), 1 (enable: thermocouple range only)

 

printing

SHIFT

disable

8.2

RANGE

 

 

range*

 

 

 

Printing

0.000 to 5.000

Max. 10 digits (minimum value to maximum

 

 

 

 

 

 

 

 

 

 

range

 

value). Excluding decimal point.

 

°C/°F

SHIFT

+

 

 

°C (°C calculation) °F (°F calculation)

 

°C

Thermocouple and resistance thermometer

8.3

calculation

(3 sec. or more)

 

 

range only

 

 

 

 

 

 

 

 

Chart speed

SHIFT

+

1

20/25 *

0001 to 0600 mm/h or 0001 to 0200 mm/min.

8.4

CHART

* RD200: 20 mm/h, RD2800: 25 mm/h

2

Other Programmable Parameters and Functions

Parameters

Key operations

Programming

Defaults

Programming ranges

Ref.

items

page

 

 

 

 

Japanese

 

Clock

SHIFT

+

CLOCK

2000.01.01.00:00 to 2099.12.31.23:59

11.1

time

 

 

 

 

 

Max. 11 digits (minimum value to maximum value),

 

 

 

 

 

 

 

 

Scale *

SHIFT

+

4

 

0.000 to 5.000

excluding decimal point. Printing range reflects to the

11.2

scale. Therefore no programming is normally

SCALE

 

 

 

 

 

 

necessary for thermocouple and resistance

 

 

 

 

 

 

 

thermometer range.

 

 

 

 

 

 

 

Programming is required in [Range/Printing range]

 

 

 

 

 

 

 

programming mode.

 

Skip *

SHIFT

+

7

Not

Channel to be deleted Clear the range/printing

11.3

RANGE

programmed

range.

 

 

 

 

 

 

 

 

 

 

 

 

Channel to be recovered Program the range/

 

 

 

 

 

 

 

printing range.

 

 

 

 

 

 

 

Programming is required in [Range/Printing range]

 

Subtract

 

+

 

 

Not

programming mode.

11.4

 

7

[Subtract printing CH] = [Reference CH] – [subtraction CH*]

printing *

SHIFT

programmed

 

RANGE

 

* Can be replaced with [reference value (with decimal

 

 

 

 

 

 

 

point)].

 

 

 

 

 

CH/output

Not

Select from CH (1 to 4), Fail* or C.End (chart end).

 

 

 

 

 

type

programmed

* Recorders with alarm output (option) only.

 

 

 

 

 

Level

Not

Up to 4 alarm points can be programmed to each

 

 

 

 

 

programmed

channel (CH 1 to 4).

 

 

 

 

 

 

 

 

 

 

 

 

 

Applicable to CH (1 to 4) selected in CH/output type

 

 

 

 

 

 

 

mode.

 

 

 

 

 

Alarm

H

H/L (absolute value high/low limits)

 

 

 

 

 

E/F (absolute value high/low limits with standby)

 

 

 

 

 

types

 

 

 

 

 

 

U/d (rate-of-change increase/decrease limits)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

b/S (differential high/low limits)

 

Alarm

 

 

8

 

 

J/k (differential high/low limits with standby)

 

SHIFT ALARM

 

 

 

11.5

 

 

Applicable to recorders with alarm output (option)

 

 

 

 

Relay No.

0

Relay No. "0" to "n" can be programmed. ("0": no

 

 

 

 

 

output), n: number of output points.

 

 

 

 

 

 

 

For recorders without alarm output (option), the

 

 

 

 

 

 

 

number is fixed as “0”.

 

 

 

 

 

Alarm

Not

Max. 5 digits including (-) sign.

 

 

 

 

 

values

programmed

Sign is not necessary when U/d, b/S or J/k is selected.

 

 

 

 

 

Measuring

Not

1 to 20, Programming required only when U/d is

 

 

 

 

 

count

programmed

selected.

 

 

 

 

 

Ref. CH

Not

1 to 4, Programming required only when b/S and J/k

 

 

 

 

 

programmed

are selected.

 

 

 

 

 

 

 

* Programming is required to each channel. For 1-pen type, programming is only for CH 1. The items without * mark are programming common to all channels. For "alarm", programming is required to each level of channels.

-34-

Page 38
Image 38
Omega RD200 List of Programming Items, Parameters to be Programmed First, Other Programmable Parameters and Functions