10. OPERATIONS

10.2 Operations at Abnormal Inputs

1 Over-range input

The following table shows displays and printings for input exceeding a printing range or a measuring range.

Measuring range:

Determined by the range No. programmed in “Range/Printing range”. See Section 8.2.

Printing range:

Trace printing range programmed in “Range/Printing range”.

Definition of input

Measuring range

(1)

(2)

Printing range

(3)

(4)

区分

Input

Display

 

Printing

Digital

 

Bar graph

Digital

Analog

 

 

 

 

Lower than

 

□□□……□□□

 

 

(1)

minimum

 

- - - - -

 

_ _ _ _ _

 

(All go off.)

 

 

measuring range*

 

 

Overshooting to

 

 

 

 

 

 

Lower than

 

□□□……□□□

 

minimum limit

(2)

minimum printing

Normal

Normal

 

 

(All go off.)

 

 

range

 

 

 

 

 

 

 

 

 

 

 

Higher than

 

■■■……■■■

 

 

(3)

maximum printing

Normal

Normal

 

 

(All light.)

 

 

range

 

 

 

Overshooting to

 

 

 

 

 

 

Higher than

_ _ _ _ _

■■■……■■■

 

maximum limit

(4)

maximum

 

+ - - - -

 

 

 

(All light.)

 

 

measuring range*

 

 

 

 

 

 

 

 

 

 

*For the digital display and printing, the measured values can be displayed and printed up to about ±10% of the span even when the values are outside the measuring range.

2

Input Disconnection

The display and printing, when the input is disconnected, differ depending on the “Burnout enable/disable” programming. However, for channels of a voltage (mV or V) range, the burnout function is “disable (non)” even when “Burnout Enable/Disable” is programmed. Program “Burnout enable/disable” to each channel. See Section 11.11.

Programming burnout

 

Display

 

Printing

Digital

 

Bar graph

Digital

 

Analog

 

 

 

Disable

Undefined

 

Undefined

Undefined

 

Undefined

(non)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Enable - downscale



 

□□□……□□□

BURN

 

Overshooting to

(doWn bUrn)

 

(All go off.)

 

minimum limit

 

 

 

 

 

 

 

 

 

 

 

Enable - upscale



 

■■■……■■■

BURN

 

Overshooting to

(UP bUrn)

 

(All lights.)

 

maximum limit

 

 

 

 

 

 

 

 

 

 

 

-54-

Page 58
Image 58
Omega RD200, RD2800 manual Operations at Abnormal Inputs, Over-range input, Input Disconnection, Definition of input