18. ADJUSTMENT

18.4 Time-axis Adjustment of Pens

When the time-axis synchronization (POC) is programmed to “on”, the gaps between the pens may change over time, resulting in error between their time-axis. This adjustment is for correction of these changes. It is recommended to adjust them once a year for maintaining the time-axis synchronization.

Adjustment Flow Chart

Operation screen

DISP

2 sec. or more

 

 

 

 

 

 

 

 

 

SHIFT

6

 

 

 

 

 

 

 

 

 

 

A - Z

[Engineering mode selection]

 

 

 

3 sec. or more

Ref.1

 

 

 

 

 

 

 

 

 

 

 

 



  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cursor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Adjustment

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

°

 

 

°

 

 

ENTRY

[Selecting adjustment]

 

 

 

 

 

F

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 



  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cursor

°

 

 

°

 

 

ENTRY

 

 

 

 

 

 

 

 

 

 

 

F

C

 

 

 

[Selecting time-axis]

 

 

 

 

 

 

 

 

 

 

 

 

 

Ref.2

 



  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cursor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Time axis

 

[Selecting an adjusting pen]

 

°F

°C

 

 

Ref.3

 

 

 

 

 

 

 

 

 

 

 



   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(1)

 

(2)

(3)

 

(4)

 

 

 

 

 

 

 

 

 

 

 

 

ENTRY (Starting adjustment )

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When (1) the reference pen prints a straight line on

 

 

0% to 100% of the chart and (2) Press

/

, the

 

 

chart is fed and then the printing of the adjusting pen

 

 

coincides with the straight line (Note)

 

 

 

 

 

 

 

 

Note) If you press

ENTRY

 

while it does not coincide

 

 

 

with the straight line, re-execute the procedure

 

 

 

from the beginning of [Selecting an adjusting

 

 

 

pen].

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENTRY (Computation of

 

 

 

 

 

 

 

 

 

 

 

 

correction data)

 

 

 

 

Temporarily storing

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

YES

Programming other channel?

NO

SHIFT

SET

 

 

END

Stored

 

Ref.4

Adjustment procedure

(1) Reference pen prints a straight line.

(3) When these two lines match, press

ENTRY .

(2) Initial position of the pen adjusted

Other pens than the pen being adjusted

Pens stand by at 0% position of scale.

Reference 1 Engineering mode types

: Memory clear

: Adjustment

: Hardware check

Reference 2 Adjustment types

: Measured value adjustment

: Shift programming for measured value

: Printing position adjustment

: Time-axis adjustment for pens

Reference 3

 

Adjusting pen selection

 

 

 

 

 

 

 

 

 

 

Display

Reference

Pen to be

 

 

pen

adjusted

 

 

 

 

 

 

 

(1) 1 P

Plotter pen

1st pen

 

 

(2) 2 P

1st pen

2nd pen

 

 

(3) 3 P

1st pen

3rdt pen

 

 

(4) 4 P

1st pen

4th pen

 

 

 

 

 

Reference 4

 

Storing

 

 

 

Store the computed correction data into memory.