8.BASIC PROGRAMMING

8.1 Programming Parameters Before Operation

Certain parameters are to be programmed for starting operation after turning on the power supply.

1 Turning On the Power supply

By turning on the power supply for the first time, the display and printing operation start with default parameters. Perform the basic programming to match your purpose before operation.

Default parameters

Range/

07 (±5 V)/0.000 to 5.000 (V)

Printing range

 

°C/°F selection

°C calculation

 

 

Chart speed

RD200: 20 mm/h

RD2800: 25 mm/h

 

 

 

2

Basic Programming Parameters

Be sure to program the following three parameters before operation.

Turning on the power supply

Programming range/printing

range*

NO

Is thermocouple or resistance

thermometer range used?

Program the following items for each channel to be used.

(1)Range No.

(2)Trace printing range

(3)Reference junction compensation (RJ): Enable or disable (only for thermocouple range)

YES

Selecting °C/°F computation

Programming chart speed

Basic programming end

Parameters to be

programmed depending on

functions

Parameters to be programmed

depending on options

Select either computing unit “°C” or “°F” to measure temperature when thermocouple or resistance thermometer range is used. The default parameter is “°C”.

Default speed: RD200 is set to 20 mm/h RD2800 is set to 25 mm/h

Default programming range: 0001 to 0600 mm/h or 0001 to 0200 mm/h

(1)Scale* (2) Skip (channel deletion and recovery) (3) Subtract printing (4) Alarms (absolute value, rate-of-change and differential) (5) Periodic data printing

(6)Clock (7) Engineering unit (8) Tag (9) Burnout enable/disable (10) Pass code/Keylock (11) Alarm dead band (12) Message (13) Time axis synchronization

(1)Alarm output (2) Communications interface (Note) (3) Remote contacts (4) Printing format (automatic range shift, compressed/expanded, zone record)

(Note) Refer to the separate instruction manual for “Communications Interface”.

*When the printing range is set, that value reflects to the “Scale” setting. For the scaling with voltage input, read the Section 11.2.

-39-