22. SPECIFICATIONS

9 Accuracy Ratings

No.

Input type

Measuring range

Reference

Accuracy

range

rating

 

 

 

01

 

-13.80 to 13.80mV

±13.8mV

 

 

 

02

DC

-27.60 to 27.60mV

±27.6mV

 

 

 

03

-69.00 to 69.00mV

±69.0mV

 

 

 

(mV)

 

 

 

04

-200.0 to 200.0mV

±200.0mV

 

 

 

 

 

 

 

05

 

-500.0 to 500.0mV

±500.0mV

±0.1% ± 1 d

06

 

-2.000 to 2.000V

±2V

 

 

 

 

07

DC

-5.000 to 5.000V

±5V

 

 

 

08

-10.00 to 10.00V

±10V

 

 

 

(V)

 

 

 

09

-20.00 to 20.00V

±20V

 

 

 

 

 

 

 

10

 

-50.00 to 50.00V

±50V

 

 

 

21

K

-200.0 to 300.0°C

±13.8mV

 

 

 

22

-200.0 to 600.0°C

±27.6mV

 

 

 

23

 

-200.0 to 1370 °C

±69.0mV

 

 

 

24

E

-200.0 to 200.0°C

±13.8mV

 

 

 

25

-200.0 to 350.0°C

±27.6mV

 

 

 

26

 

-200 to 900°C

±69.0mV

 

 

 

27

J

-200.0 to 250.0°C

±13.8mV

 

 

 

28

-200.0 to 500.0°C

±27.6mV

±0.1% ± 1 d

29

 

-200 to 1200°C

±69.0mV

 

 

 

 

30

T

-200.0 to 250.0°C

±13.8mV

 

 

 

31

-200.0 to 400.0°C

±27.6mV

 

 

 

 

 

 

 

32

R

0 to 1200 °C

±13.8mV

 

 

 

33

0 to 1760 °C

±27.6mV

 

 

 

 

 

 

 

34

S

0 to 1300 °C

±13.8mV

 

 

 

35

0 to 1760 °C

±27.6mV

 

 

 

 

 

 

 

36

B

0 to 1820 °C

±13.8mV

 

 

 

37

N

-200.0 to 400.0°C

±13.8mV

 

 

 

38

-200.0 to 750.0°C

±27.6mV

±0.15% ± 1 d

39

 

-200 to 1300 °C

±69.0mV

 

 

 

 

40

WWRe26

0 to 2315°C

±69.0mV

 

 

 

41

WRe5

0 to 2315°C

±69.0mV

 

 

 

WRe26

 

 

 

43

PtRh40-

0 to 1888°C

±13.8mV

 

 

 

 

PtRh20

 

 

±0.2% ± 1 d

44

NiMo-Ni

-50.0 to 290.0°C

±13.8mV

 

 

 

45

-50.0 to 600.0°C

±27.6mV

 

 

 

46

 

-50 to 1310 °C

±69.0mV

 

 

 

47

CR-AuFe

0.0 to 280.0K

±13.8mV

 

 

 

48

Platinel ll

0.0 to 350.0°C

±13.8mV

 

 

 

49

0.0 to 650.0°C

±27.6mV

 

 

 

50

 

0 to 1395°C

±69.0mV

±0.15% ± 1 d

51

U

-200.0 to 250.0°C

±13.8mV

 

 

 

52

-200.0 to 500.0°C

±27.6mV

 

 

 

53

 

-200.0 to 600.0°C

±69.0mV

 

 

 

54

L

-200.0 to 250.0°C

±13.8mV

±0.1% ± 1 d

55

-200.0 to 500.0°C

±27.6mV

56

 

-200 to 900°C

±69.0mV

 

 

 

70

Pt100

-140.0 to150.0°C

160Ω

±0.15% ± 1 d

71

-200.0 to 300.0°C

220Ω

 

 

 

(JIS’97)

±

±

1 d

 

 

 

72

 

-200.0 to 850.0°C

400Ω

0.1%

 

 

 

 

 

73

Pt100

-140.0 to 150.0°C

160Ω

±0.15% ± 1 d

74

QPt100

-200.0 to 300.0°C

220Ω

±0.1% ± 1 d

75

JIS’89

-200.0 to 649.0°C

400Ω

 

 

 

76

JPt100

-140.0 to 150.0°C

160Ω

±0.15% ± 1 d

77

-200.0 to 300.0°C

220Ω

±0.1% ± 1 d

78

 

-200.0 to 649.0°C

400Ω

 

 

 

 

No.

Input type

Measuring range

Reference

Accuracy

range

rating

 

 

 

79

Pt50

-200.0 to 649.0°C

220Ω

±0.1% ± 1 d

80

Pt-Co

4.0 to 374.0K

220Ω

±0.15% ± 1 d

Note) “± 1 d” at accuracy rating stands for ± 1 digit.

Note) Thermocouple inputs do not include reference junction

 

compensation accuracy.

 

 

 

The Indication equivalent to 200 V may vary under the

 

test environment of EMC directive.

 

Exceptions to accuracy rating

Input type

Measuring range

Accuracy rating

K, E, J, T, L

-200 to 0°C

±0.2% ± 1 digit

R, S

0 to 400°C

 

B

0 to 400°C

Not specified

400 to 800°C

±0.15% ± 1 digit

 

N, U

-200 to 0°C

±0.3% ± 1 digit

W-WRe26

0 to 100°C

±4% ± 1 digit

100 to 300°C

±0.5 % ± 1 digit

 

PtRh40-PtRh20

0 to 300°C

±1.5% ± 1 digit

300 to 800°C

±0.8% ± 1 digit

 

CR-AuFe

0 to 20K

±0.5% ± 1 digit

20 to 50K

±0.3% ± 1 digit

 

Pt100 (’97)

700 to 850°C

±0.15% ± 1 digit

Pt-Co

4 to 50K

±0.3% ± 1 digit

Note) Thermocouple input is converted into the reference range.