13. REMOTE CONTACTS
13.3 Programming 3 Chart Speeds
The “3 chart speeds” programming screen appears only in the remote contacts (option).
Press SHIFT and simultaneously to display the “3 chart speeds” programming screen. Program 3 chart speeds (CS1 to 3) and select the desired speed with a contact signal. See section 13.1 2 .
3
Programming Mode
| | |
Unit for the Cursor Speed | Chart speed | |
speed | No. | (4 digits programming) |
|
|
4 Programming Flow Chart
<Example> From 20 mm/h to 40 mm/h for CS2 |
|
|
| ||||
|
| Operation screen |
| DISP | |||
|
|
|
|
| |||
[Check] |
| SHIFT | + | 1 |
| Ref. 1 | |
| CHART | ||||||
| |
|
|
|
| ||
Speed unit | Speed No. |
|
|
|
| ||
|
| SET |
|
|
|
| |
|
| END |
|
|
| Ref. 2 | |
[To be ready for programming] |
|
|
| ||||
|
|
|
| ||||
| |
|
|
|
| ||
| Cursor |
|
| → |
| ||
[Selecting CS No.] | / | % | °F | ||||
| |||||||
|
|
|
|
| |||
| |
|
|
|
| ||
| Speed No. | 0 | 0 |
| 4 | 0 | |
[Selecting speed] | CLEAR | CLEAR | SCALE | CLEAR | |||
|
|
|
|
| |||
| |
|
|
|
| ||
|
| ENTRY |
|
|
|
| |
YES | Error display appears? |
|
|
| |||
|
|
|
| ||||
Any key other |
|
|
|
|
| ||
than SHIFT | NO |
|
|
|
| ||
|
| Temporarily stored |
|
|
|
| |
YES | Programming other Nos.? |
|
|
| |||
|
|
|
|
|
| ||
|
| NO |
|
|
|
| |
|
| SHIFT | + | SET |
|
| |
|
|
|
| END |
|
| |
|
| Stored | Ref. 3 |
| |||
|
|
|
| -90- |
Defaults parameters and
programming range
Default: 20 (RD200), 25 (RD2800)
Programming range: 0001 to 0600 (mm/h) or 0001 to 0200 (mm/min.) 0000mm/h can be set for RD2800. When 0000mm/h is set, chart is fed at the speed of 12.5mm/h.
If the chart speed is Remarks 1 programmed at 151 mm/h
or faster;
All printings except time line, digital data printing and programming change mark will not be executed. See Section 10.1.
Remarks 2 | Influence to periodic | |
data printing | ||
|
If the chart speed is changed, the parameters programmed in the [Periodic data printing] (Section 11.7) is cleared.
Reference 1 |
| Other speed No. check | ||
|
|
| ||
Press | / | or % to change CS | ||
No. Chart speeds can be checked. |
| |||
|
|
|
| |
|
| Changing the speed |
| |
Reference 2 |
| unit (h/min.) |
|
|
Place the cursor on the character |
| |||
display (left digit) and press | / | or |
%to switch [M] and [H] alternately.
Reference 3 Storing
Store the [temporarily stored] parameters into memory. (Programming change mark is printed.)