11. OTHER PROGRAMMING

11.11 Burnout

The channel where the temperature range is programmed becomes effective. Press SHIFT and % simultaneously in the operation screen for 3 seconds or more to display the [Burnout] programming screen. This programming is necessary for each channel.

1 Burnout

If a sensor (thermocouple or resistance thermometer) is disconnected, trace printing overshoots maximum or minimum limit.

The default parameter of burnout is “disable” () in all channels.

Burnout selection menu

Burnout can be selected for each channel out of 3

types shown below.

: Burnout disable

(Non)

 (UP burn)

: Up scale burnout

 (down burn)

: Down scale burnout

2 Programmed Flow Chart

<Example> From burnout “disable” to “up scale burnout”

 

Operation screen

 

DISP

 

 

 

( SHIFT

%

[Check]

 

 

3 sec. or more

Ref. 2

 

 

 

 

 

 



 

 

 

 

 

Channel

 

SET

 

 

 

 

 

 

 

 

 

[To be ready for programming]

END

 

 

 

 

 

 

 



 

 

 

 

 

 

Cursor

 

2

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

[Specifying channel]

 

 

 

 

 



 

 

 

 

 

 

Cursor

 

 

Press several times.

[Selecting burnout]

 

/

 

 

 

 

 





 

 

 

 

Cursor

 

 

 

 

 

 

 

 

ENTRY

 

 

 

 

Temporarily stored

 

 

 

YES

Programming other channels?

 

 

 

 

 

NO

 

 

 

 

 

 

 

SHIFT

SET

 

 

 

 

END

 

 

 

Stored

 

Ref. 3

 

 

 

 

 

 

-75-

” programming on

Remarks 1 parallel operation

A parallel connection of a thermocouple with other instrument causes a problem. Select () if your recorder uses the thermocouple being connected with other instrument.

Remarks 2 Voltage range is ineffective

Even if the burnout is programmed to a channel where the voltage range (No. 01 to 07) is programmed, this programming becomes ineffective (burnout disable).

Reference 1 Burnout operation

For the burnout operation (display and printing), read Section 10.2.

Reference 2 Other channels check

Press

/

or

%

to change

channels. The programmed burnout of

other channels can be checked.

 

 

 

 

Reference 3

Storing

 

 

 

Store

the

[temporarily

stored]

parameters

into

 

memory.

(Programming

change

mark is

printed.)

 

 

 

 

 

Page 79
Image 79
Omega RD2800, RD200 manual Burnout, ,  ,   