14. PRINTING FORMAT

14.2Programming Compressed/Expanded Printing

This programming display only appears for the printing format (option). This programming is to print the specified range in the [compressed/expanded] printing mode.

1 Compressed/Expanded printing

A specified range in the trace printing range can be printed in the compressed or expanded printing mode.

(1)This programming can be executed for each channel.

(2)The printing range can be programmed optionally irrespective of the programmed range in the [Range/Printing range].

(3)Up to 2 break points can be programmed.

2 Programming Mode

1) Screen 1 (Printing format check)

<Example> Compressed and expanded programming

<Range No. 23 (K thermocouple)>

Measuring value

 

-100

 

600

800

1200

 

 

 

 

 

 

 

 

 

 

 

 

-200

Com-pre

Expanded

Com-pre 1370

Printing range

ssed

 

ssed

 

 

 

 

 

 

 

 

 

 

 

(1)

 

 

 

(1)

 

 

 

 

 

 

 

 

 

 

 

 

Printing

 

 

 

 

 

 

 

 

 

 

position

0%

20%

 

 

 

80%

100%

 

 

 

(No. 1 break point)

 

(No. 2 break point)

 

(2)

(3)

 

 

 

(4)

(5)

 

Scale

 

600

 

 

 

800

1200

 

 

-100

 

 

 

 

Mark

is the programming point and No. (1) to (5) shows the

order of programming. By pressing

ENTRY after the

programming ends, the next programming screen appears. (Program No.1 break point and No.2 break point by 2 digits. <Example> 5% 05)

3)Screen 3 (Programming break point printing position)

  



 

Standard

Automatic

Zone

 

range shift

Compressed/expanded

 

 



 

 

 

 

 

 

Programming

No. 1

Programming No. 2

break point

break point (Note)

 

 

 

Dots appear at the selected format.

By pressing SET , the dots disappear and the

END

cursor appears instead.

2) Screen 2 (Programming channel)

Programming mode of compressed/expanded

 

Cursor Channel

No. 1 break

No. 2 break

 

point

point

By selecting compressed/expanded ( ), the cursor appears at the channel programming position.

(1)Program each break point printing position (%) in the range of 10 to 90.

(2)When all programming are completed, return to

the above display and execute [Storing].

(Note) No.1 break point should be lower than No.2 break point. When no programming for the No.2 break point is required, leave it as spaces.

4)Screen 4 (Scale programming of each printing position)

Printing position (%)

 

Programming value

(Max. 5 digits by left-justify)

(1)When programming a scale, the printing position advances to the next position. Program scales up to the 100% position.

(2)After completing the programming for 100%,

press ENTRY to return to Screen 3.

-94-