22. SPECIFICATIONS
1 Input Specifications
Number of measuring points: 1, 2, 3 and 4 points
Input signals: DC voltage (mV) …±13.8, ±27.6, ±69.0, ±200
±500
DC voltage (V)…±2, ±5, ±10, ±20, ±50 (Built- in voltage divider for ±5 V or more)
DC current…Available by adding shunt resistor (option)
Thermocouple…B, R, S, K, E, J, T, N,
Resistance thermometer…Pt100 (1),
Pt100 (2), JPt100, Pt50,
range No. list)
Scale: Program minimum and maximum values with key operation.
Programming
Measuring range: Refer to the list of range No. in Section
Minimum printing range: Refer to the following table
| DC |
| 1/5 or higher than the measuring range. | ||
| voltage |
| |||
|
|
|
|
| |
|
|
| 2/5 or higher than the measuring range when | ||
| Thermo- |
| converted into electromotive force. | ||
|
| (In the case of range Nos. with minus | |||
| couple |
| measuring ranges, a temperature of 0°C or | ||
|
|
| higher is specified as the | ||
|
|
| measuring value.) |
| |
| Resistance |
| Pt100 | 100°C span or more | |
| thermometer |
| Pt50 | 200°C span or more | |
| Accuracy rating: Refer to item | 6 | (at reference operating | ||
|
|
| condition). |
|
|
Temperature | drift: ±0.01% full | scale/°C (converted into |
reference range)
Measuring cycle: About 100 msec (CH1 to CH4 )
Reference junction compensation: Inside (enable) or external (disable) is selectable.
Reference junction compensation accuracy:
K, E, J, T, N and Platinel II…±0.5°C or less Other than the above*… ±0.1°C or less *Except
Input resolution: Approx. 1/56000 (converted into reference range)
Burnout: Effective input… Thermocouple, resistance thermometer
Programming range …Selection from
Allowable signal source resistance:
DC voltage inputs, thermocouple inputs
…1kΩ or less*
Resistance thermometer inputs
…10Ω or less
* Without burnout
Input resistance: Thermocouple input …About 8MΩ DC voltage inputs (range of ±2 V or less)
…About 8MΩ
DC voltage inputs (range of ±5 V or less) …About 1MΩ
Maximum input voltage
DC voltage | ±10 V DC or less (range of ±2 V or less) | |
inputs, | ||
±60 V DC or less (range of ±5 V or less) | ||
thermocouple | ||
| ||
Resistance | ±6 V DC or less | |
thermometer | ||
|
Input correction: Shift programming for each channel
Input filter: Program primary delay filter
1 to 10 sec. and “0” (none) for each channel. Maximum common mode voltage: 30VAC
Common mode rejection ratio: 140 dB or more (50/60 Hz)
Series mode rejection ratio: 50 dB or more (50/60 Hz)
Terminal board: Detachable and removable for wiring
2 Printing
Printing accuracy: Measuring accuracy + ±.0.3% of printing range (under the reference operation condition)
Printing dead and: 0.2% of printing range
Printing system:
Trace printing… Disposable cartridge pen Digital data printing …. Plotter pen
Step response: RD200 … 1 sec. or less (90% response) RD2800 … 1.5 sec. or less (90% response)
Print color: Trace printing ….1st pen (Red), 2nd pen (Green),
3rd pen (Blue), 4th pen (Brown) Digital data printing … Purple
<<Digital printing>>
cPeriodic | : Printing of time, channel No., measured |
data printing | values and engineering unit with |
| programmed interval |
dDigital data | : Printing of time, channel No., measured |
printing | values and engineering unit with trace |
eYear, Month, | printing on demand |
: Printing of year, month, day and time when | |
day, time, time | turning on the power. Printing time line and |
line | time every on the hour. Printing year, moth |
| and day at 00:00 and every specific interval |
fChannel No., | : Printing of scale, channel No. and tag (only if |
scale, tag ゙ | they are programmed) with every specific |
| interval |
gChart speed | : Printing of chart speed with every specific |
| interval |
hAlarm | : Printing of time, channel No., alarm type and |
Activation/ | alarm level when an alarm activates. Printing |
Reset | time, channel No. and alarm level when the |
| alarm is reset. Memory volume is max. 48 |
| data |
iProgramming | : Marking when changed parameters are |
change mark | stored |
jPOC mark | : Marking when |
| programmed |
kList printing | : Printing of parameters (list 1 or 2) on demand |
| List 1: Chart speed, range/printing range, |
| scale, subtract printing, periodic data |
| printing, printing format and alarm |
| List 2: Chart speed (3 speeds), alarm output, |
| |
| contacts and math expression |
| List 3: All parameters (List 1 + List 2) |
-126-