OPERATION
Options Window
350-9047 v3.0.1 DGy Model 201x User’s Guide 57
. . . . .
Update Period
The internal clock is synchronized to the currently
selected reference signal (NTS or IRIG) periodically.
Adjust the length of time between updates by defining
the update period in the entry box provided. The default
period is 2 hours but the rate may be set from a
minimum update period of 30 minutes to a maximum
update period of 24 hours; entries may be made in
increments of 30 minutes. Having a short update period
may be required when using multiple DGy 201x
recorders in a time-synchronized application.
REMOTE SHUTDOWN
Click the Shutdown button to remotely power off the
DGy 201x. The recorder will go through the same
shutdown procedure as when the front panel power
switch is used. Follow the on-screen prompts to confirm
or cancel shutdown. Shutdown will take 10 to 15
seconds to complete. During shutdown, the power
indicator on the DGy will remain lit, but the system will
be inoperable. When shutdown is complete, the WCP
may be closed.
Important Do not remove the power cord from the rear of the
DGy until the shutdown process is complete. Failure
to wait for the normal shutdown procedure to execute
may cause loss of data.