X20 Sprayer Reference Manual

Rate

The Rate display is filtered to prevent the numbers

 

from continually changing. If the achieved rate is

 

within 5% of the target rate, the display will only

 

show the target rate, but the system will still control

 

to the target rate. If an error of less than 5% persists,

 

the error is reported as a rate alarm after approxi-

 

mately 1 minute. When first turning the Sprayer on,

 

the system should control to rate within 10 seconds,

 

subject to the position of the Regulator Valve when it

 

was last turned off. If the 'Close Valve When Off' is

 

selected with a slow valve, ie, 20 second type, it may

 

take longer, but this is because the option is selected

 

when it shouldn't be. Because of the filtering, the

 

display may take up to 30 seconds to stabilize if there

 

is a long adjustment required to change the rate with

 

a slow valve. This can result in a false Rate alarm

 

where the actual rate displayed is still catching up to

 

the target rate. The Sprayer is usually controlling cor-

 

rectly, which can be checked by checking the pressure

 

reading, it is just the display that is sluggish.

 

The alarm default is set to 'Repeat While Present'

 

and as the alarm is critical, needs to sound while the

 

condition is present. As applies to all alarms, they can

 

be changed to different cycle patterns, with differing

 

delays.

Tank Empty

This alarm simply states that the calculated volume

 

has reached zero. If there is still spray in the tank,

 

the system will still operate giving the volume used

 

still accumulating, and the tank volume as a negative

 

figure. The default setting is one alarm cycle only, but

 

other alarms will sound when the tank realy runs out.

18-20

www.topconpa.com

Page 246
Image 246
Topcon A2677 manual Figure. The default setting is one alarm cycle only, but