Controlling the E5250A

Switch Control Basics

E5252A Channel List

This section describes how to define the channel_list specifically for the E5252A.

For the basic syntax of the channel_list, see “Channel List and Channel

Configuration” on page 5-6.

 

Normal Configuration Mode

 

 

 

 

 

To specify a channel, you use five digits as shown in Table 5-5.

 

Table 5-5

Specifying E5252A Channels in Normal Config. Mode

 

 

 

 

 

 

 

 

 

 

 

One Channel: 5 digits

 

 

Slot No.

 

 

 

 

 

 

 

 

Channel Number: 4 digits

 

(E5252A is

Card Number

 

 

 

 

 

 

 

installed)

 

 

 

 

 

(first 1 digit)

 

Input Port No.

 

Output Port No.

 

 

 

 

 

 

 

 

(next 2 digits)

 

(last 2 digits)

 

 

 

 

 

 

 

 

1

1

 

01 to 10

 

01 to 12

 

 

 

 

 

 

 

 

2

2

 

01 to 10

 

01 to 12

 

 

 

 

 

 

 

 

3

3

 

01 to 10

 

01 to 12

 

 

 

 

 

 

 

 

4

4

 

01 to 10

 

01 to 12

 

 

 

 

 

 

 

 

Card Number:

One digit: 1 to 4. E5250A slot number in which card is

 

 

installed.

 

Channel Number: Four digits: XXYY.

 

 

• XX: input port number (01 to 10).

 

 

• YY: output port number (01 to 12).

 

 

Channel Number assignment is shown in Figure 5-4.

 

For the following two cases, three ports share one input path. Do not specify two of

NOTE

 

the input ports at same time.

• SMU5, HF1, and CV1 (Input Port No. 05, 07, and 09)

• SMU6, HF2, and CV2 (Input Port No. 06, 08, and 10)

For example, if you specify input port 05, then you cannot specify input ports 07 and 09. You can specify 05 and 06 at the same time.