Executing Self-Test and Leak Test

 

This chapter describes how to execute the Self-Test and the Leak Test.

 

The Self-Test consists of three tests that check the operation of Agilent E5250A

 

firmware, LEDs, and key that are on the E5250A front panel, and the relays on the

 

cards installed in the E5250A.

 

Self-Test should be executed every day as an operation check before using the

 

E5250A.

 

The Leak Test checks leakage current of cards installed in the E5250A.

 

Leak Test can be executed by using Agilent 4155/4156 Semiconductor Parameter

 

Analyzer with the Self-Test Utility program that is on the E5250A Program Disk

 

furnished with the E5250A.

 

Performance Verification

NOTE

 

The Self-Test and the Leak Test cannot be used to verify the specifications of the

 

E5250A, E5252A, and E5255A.

 

To verify the specifications, return the E5250A with the cards installed to your

 

nearest Agilent Technologies Service Center. The performance verification should

 

be done by service personnel who are trained to service the E5250A, and should be

 

performed once a year.

 

• “Executing Self-Test” introduces the test items of the Self-Test and describes

 

 

how to execute the Self-Test.

 

• “Using the Self-Test Utility” describes how to execute the Self-Test and the

 

Leak Test interactively from Agilent 4155/4156 semiconductor parameter

 

analyzer.

3-2

Agilent E5250A User’s Guide, Edition 9

Page 52
Image 52