Command Reference *SRE

*SRE

This command sets the Service Request "Enable" Register bits. 1 enables, 0 masks.

Syntax

*SRE enable_number

 

 

Parameter

Explanation

 

 

 

 

 

 

enable_number

decimal integer (that is the sum of the binary-weighted values

 

 

 

 

for the desired bits), hexadecimal, octal, or binary value

 

 

 

 

 

Query response

enable_number <newline><^END>

Semantics

The Service Request "Enable" Register consists of 8 bits: Bit0 to Bit7. Bit6 is not

 

defined, and is always 0. The Service Request "Enable" Register determines which

 

bits of the Status Byte Register are enabled.

 

The status of the enabled bits are ORed together, and the result of OR is output to

 

bit6 (Master Summary Status bit) of Status Byte Register. For details, refer to

 

“Status Reporting Structure” on page 7-48.

 

The following table shows the bits of the Status Byte Register, and the

 

binary-weighted decimal value of each bit.

 

 

 

 

 

 

 

bit

binary-

description

 

 

weight

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

1

not used

 

 

 

 

 

 

 

 

1

 

2

not used

 

 

 

 

 

 

 

 

2

 

4

not used

 

 

 

 

 

 

 

 

3

 

8

not used

 

 

 

 

 

 

 

 

4

 

16

MAV (Message Available summary-message)

 

 

 

 

 

 

 

 

5

 

32

ESB (Event Status Bit)

 

 

 

 

 

 

 

 

6

 

64

MSS (Master Summary Status)

 

 

 

 

 

 

 

 

7

 

128

not used

 

 

 

 

 

 

 