CAUTION

Installation

Requirements

Requirements

This section describes the following requirements for the E5250A.

“Power Requirements”

“Power Cable”

“Operating Environment”

“Storage and Shipping Environment”

Power Requirements

Before applying ac line power to the E5250A, ensure that the correct power cable is used.

The E5250A can operate from any single-phase ac power source supplying 100 to 240 V ±10 % in the frequency range from 47 to 63 Hz. The maximum power consumption is 100 VA. For details, refer to Chapter 10, “Specifications.”

Power Cable

In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument frame. The type of power cable shipped with each instrument depends on the country of destination. Refer to the following table for the part numbers of the power cables available.

If the plug on the cable does not fit the power outlet, or the cable is to be attached to a terminal block, cut the cable at the plug end and re-wire it. This work should be performed by a qualified electrician all local electrical codes being strictly observed.

The color coding used in the cable will depend on the cable supplied. If a new plug is to be connected, it must meet local safety requirements and include the following features:

Adequate load-carrying capacity. (refer to Chapter 10, “Specifications.”)

Ground connection.

Cable clamp.

2-4

Agilent E5250A User’s Guide, Edition 9

Page 36