Programming the E5250A

Using the VFP Data Upload Library

Connect_vfp subprogram

Makes relay connections based on data from the specified Setup Memory. Data from all 24 Setup Memories was loaded by the Load_vfp_data subprogram.

Syntax:

Connect_vfp(Setup_data$,Setup_num)

Parameters:

Setup_data$:

Data string for internal use in the library. Must declare 14000

 

 

byte data size.

 

Setup_num:

Setup Memory number: 1 to 24.

Example:

CALL Connect_vfp(Mat48_data$,1)

FNGet_vfp_comment$ subprogram

Returns comment from the specified Setup Memory. Comments for all 24 Setup

Memories were loaded by the Load_vfp_data subprogram.

Syntax:

FNGet_vfp_comment$(Setup_data$,Setup_num)

Parameters:

Setup_data$:

Data string for internal use in the library. Must declare 14000

 

 

byte data size.

 

Setup_num:

Setup Memory number from which to get comment: 1 to 24.

Example:

Comment$=FNGet_vfp_comment$(Mat48_data$,1)

 

DISP "Comment=";Comment$

6-24

Agilent E5250A User’s Guide, Edition 9

Page 160
Image 160
Agilent Technologies Agilent E5250A manual Connectvfp subprogram, FNGetvfpcomment$ subprogram