Controlling the E5250A

Switch Control Basics

Auto Configuration Mode

To use this mode, the E5255A must be installed in the E5250A slots continuously from slot1 (slots 1,2; or slots 1,2,3; or slots 1,2,3,4). To specify a channel, use five digits as shown in Table 5-8,

Table 5-8

Specifying E5255A Channels in Auto Config. Mode

 

 

 

 

 

 

 

 

IV Port Input

 

One Channel: 5 digits

 

 

 

 

 

 

 

Connection

Card Number

Channel Number: 4 digits

 

(SMU Input

 

 

 

 

Input Port No.

 

Output Port No.

 

Port No.)

(first 1 digit)

 

 

 

 

(next 2 digits)

 

(last 2 digits)

 

to SMU1 (1)

0

IV Input: 01

 

01 to 96

 

 

 

BIAS Input: 11

 

 

 

 

 

 

 

 

 

to SMU2 (2)

0

IV Input: 02

 

01 to 96

 

 

 

BIAS Input: 12

 

 

 

 

 

 

 

 

 

to SMU3 (3)

0

IV Input: 03

 

01 to 96

 

 

 

BIAS Input: 13

 

 

 

 

 

 

 

 

 

to SMU4 (4)

0

IV Input: 04

 

01 to 96

 

 

 

BIAS Input: 14

 

 

 

 

 

 

 

 

 

to SMU5 (5)

0

IV Input: 05

 

01 to 96

 

 

 

BIAS Input: 15

 

 

 

 

 

 

 

 

 

to SMU6 (6)

0

IV Input: 06

 

01 to 96

 

 

 

BIAS Input: 16

 

 

 

 

 

 

 

 

 

Card Number:

One digit: always 0.

 

 

Channel Number: Four digits: XXYY.

 

XX: input port number.

01, 02, 03, 04, 05, or 06 for IV Input Port. 11, 12, 13, 14, 15, or 16 for BIAS Input Port.

YY: output port number.

Minimum output port number is 01.Maximum output port number is the total number of output ports for the E5255A blocks connected to the same input port.

For each configured multiplexer, output port number 01 is assigned to first output port of smallest block number on card in the lowest slot number.

Agilent E5250A User’s Guide, Edition 9

5-21