Executing Self-Test and Leak Test

 

 

 

 

Executing Self-Test

Table 3-2

SCPI Commands for Self-Test

 

 

 

 

 

 

 

 

Test Item

Description

SCPI Command

 

 

 

 

 

 

 

Controller

Executes Controller Test,

:DIAG:TEST:FRAM:EXEC? CONT

 

 

Test

then returns test result.

 

 

 

 

 

 

 

 

 

Returns test result.

:DIAG:TEST:FRAM:STAT? CONT

 

 

 

Clears test result.

:DIAG:TEST:FRAM:CLE CONT

 

 

Front Panel

Executes I/F Test, then

:DIAG:TEST:FRAM:EXEC? FPAN

 

 

Interface

returns test result.

 

 

 

(I/F) Test

 

 

 

 

Returns test result.

:DIAG:TEST:FRAM:STAT? FPAN

 

 

 

 

 

 

Clears test result.

:DIAG:TEST:FRAM:CLE FPAN

 

 

Relay Test

Executes Relay Test, then

:DIAG:TEST:CARD:EXEC? ALL

 

 

 

returns test result.

or Card No. (1, 2, 3, or 4)

 

 

 

Returns test result.

:DIAG:TEST:CARD:STAT? ALL

 

 

 

 

or Card No. (1, 2, 3, or 4)

 

 

 

Clears test result.

:DIAG:TEST:CARD:CLE ALL or

 

 

 

 

Card No. (1, 2, 3, or 4)

 

 

Front Panel Interface Test

 

NOTE

 

 

 

After the :DIAG:TEST:FRAM:EXEC? FPAN command is executed, the Remote,

System Fail, Fail, and Local/Self Test LEDs on the E5250A front panel blink for about 10 seconds. If an LED does not blink, the LED is defective.

Confirm that the four LEDs are blinking, then press the Local/Self Test key on the

E5250A front panel within 10 seconds. The LEDs should stop blinking.

If you do not press the Local/Self Test key within 10 seconds, the E5250A returns a fail result for the Front Panel Interface Test, which turns on the Fail LED.

If you pressed Local/Self Test key within 10 seconds, and the Fail LED still turns on, it means the Local/Self Test key is defective.

If you find any problem, contact your nearest Agilent Technologies Service Center.