Controlling the E5250A

Using Virtual Front Panel Utility

Softkey Menu 1:

 

 

Softkey

Description

 

 

OPEN/

Changes the relay state (open or close) of crosspoint selected by the

CLOSE

cursor. The softkey toggles the state between open and close. The

 

E5250A settings are changed in real time.

 

If bias mode is ON and cursor is in the row of input bias port, this

 

softkey changes the relay state; open, close, and bias disabled.

 

. means open.

 

@ means close.

 

− means bias disabled.

 

When bias mode is ON, if you disconnect a bias enabled output port

 

from a non-bias input port, the output port is automatically

 

connected to the input bias port.

 

 

UP

Moves the cursor upward.

 

 

LEFT

Moves the cursor to the left.

 

 

RIGHT

Moves the cursor to the right.

 

 

DOWN

Moves the cursor downward.

 

 

MORE

Displays more softkeys. The displayed softkeys depend on whether

 

cursor selects input Bias Port and Bias Mode is ON. See Softkey

 

Menu 2 and Menu 3 on next pages.

 

 

RETURN

Displays the previous level of the softkey menu, shown previously

 

in "To Modify the Settings in Current Setup Memory".

 

 

NOTE

Moving the Cursor

 

Move the cursor by using the UP, DOWN, RIGHT, LEFT softkeys. Do not use

 

arrow keys on the 4155/4156 front panel.

 

 