Command Reference *RST

*RST

This command performs an instrument reset. Status after *RST is shown below:

Channel Configuration:

Normal Config Mode

Connection Rule:

Free

Connection Sequence:

Break Before Make

Bias Mode:

Off

 

Bias Port:

E5252A: 10

 

E5255A: None

Bias Enabled Channels:

All channels (output ports) are bias enabled

Couple Port Mode:

Off

 

Couple Port:

No Couple Port

Card Channel Status:

All relays open

Self-Test Result:

No Change

Syntax*RST

Example OUTPUT @Hp5250;"*RST"

NOTE

This command does not change the self-test result.

 

For the power-on settings, see :SYSTEM:CPON.

 

 

7-10

Agilent E5250A User’s Guide, Edition 9

Page 178