Agilent E5250A User’s Guide, Edition 9 12- 3
SCPI Command Summary
Command Description
:DIAG:TEST:CARD:CLE card_number Clears relay test result (pass/fail) of specified card.
card_number: 1, 2, 3, 4, or ALL
:DIAG:TEST:CARD[:EXEC]? card_number Executes relay test, then returns result:
1: fail card exists, 0: pass
card_number: 1, 2, 3, 4, or ALL
:DIAG:TEST:CARD:STAT? card_number Returns most recent relay test result:
1: fail, 0: pass, -1: not tested.
card_number: 1, 2, 3, 4
:DIAG:TEST:FRAM:CLE unit Clears specified test result.
unit: CONT (controller test), or FPAN (front panel
interface test)
:DIAG:TEST:FRAM[:EXEC]? unit Executes specified test, then returns test result:
1: fail, 0: pass.
unit: CONT (controller test), or FPAN (front panel
interface test)
:DIAG:TEST:FRAM:STAT? unit Returns most recent test result of specified test:
1: fail, 0: pass, -1: not tested.
unit: CONT: Controller test FPAN: Front panel
interface test
[:ROUT]:BIAS:CHAN:DIS:CARD card_number Bias disables all channels of specified card.
card_number:
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
[:ROUT]:BIAS:CHAN:DIS[:LIST] (@channel_list)Bias disables the specified channels.
channel_list: Channels to bias disable.
[:ROUT]:BIAS:CHAN:DIS[:LIST]? (@channel_list)Returns whether specified channels are bias
disabled. 1: disabled, 0: enabled.
channel_list: Channels to check.