SCPI Command Summary

Command

Description

 

 

:DIAG:TEST:CARD:CLE card_number

Clears relay test result (pass/fail) of specified card.

 

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD[:EXEC]? card_number

Executes relay test, then returns result:

 

1: fail card exists, 0: pass

 

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD:STAT? card_number

Returns most recent relay test result:

 

1: fail, 0: pass, -1: not tested.

 

card_number: 1, 2, 3, 4

:DIAG:TEST:FRAM:CLE unit

Clears specified test result.

 

unit: CONT (controller test), or FPAN (front panel

 

interface test)

 

 

:DIAG:TEST:FRAM[:EXEC]? unit

Executes specified test, then returns test result:

 

1: fail, 0: pass.

 

unit: CONT (controller test), or FPAN (front panel

 

interface test)

 

 

:DIAG:TEST:FRAM:STAT? unit

Returns most recent test result of specified test:

 

1: fail, 0: pass, -1: not tested.

 

unit: CONT: Controller test FPAN: Front panel

 

interface test

 

 

[:ROUT]:BIAS:CHAN:DIS:CARD card_number

Bias disables all channels of specified card.

 

card_number:

 

For Auto Config mode: 0 or ALL

 

For Normal Config mode: 1, 2, 3, 4, or ALL

[:ROUT]:BIAS:CHAN:DIS[:LIST] (@channel_list)

Bias disables the specified channels.

 

channel_list: Channels to bias disable.

 

 

[:ROUT]:BIAS:CHAN:DIS[:LIST]? (@channel_list)

Returns whether specified channels are bias

 

disabled. 1: disabled, 0: enabled.

 

channel_list: Channels to check.

 

 