Error Messages

This chapter lists and describes the error messages for Agilent E5250A. An error message consists of an error number and message.

There are two types of error messages:

Standard SCPI Error Messages

Negative error numbers (Command Error, Execution Error, Device-Dependent Error, and Query Error) are standard SCPI errors.

E5250A Specific Error Messages

Positive error numbers are the E5250A specific errors.

Error messages are classified by error number as listed in the following table:

Error Range

Error Category

 

 

0

No error

 

 

-100 to -199

Command Error

-200 to -299

Execution Error

-300 to -399

Device-Dependent Error

-400 to -499

Query Error

100 to 32767

E5250A Specific Error

 

 

When an error occurs, the corresponding bit is set in the Standard Event Status Register (see Chapter 7):

Error Category

Standard Event

Status Register Bit

 

 

 

Command Error

bit5

 

 

Execution Error

bit4

 

 

Device-Dependent Error

bit3

 

 

Query Error

bit2

 

 

E5250A Specific Error

bit3

 

 

Also, the error number and message are placed in the error queue, which can be read by the :SYSTem:ERRor? query command. The error queue is cleared by the common command *CLS, and when power is turned on. For these commands, see Chapter 7.

11-2

Agilent E5250A User’s Guide, Edition 9

Page 306