Executing Sample Programs

HCI Measurement Program

Modifying the Program

This section explains how to modify the SP_MUX_M sample program and PARAM.MES setup file.

“To change the GPIB address”

“To change the 4155/4156 setup file used”

“To change Vtci extraction condition”

“To change Idlin and Gmmax extraction condition”

“To change the number of devices”

“To change the stress times”

“To change the test limits”

“To specify the data displayed on the 4155/4156 screen”

“To change the timing for saving the test data”

To change the GPIB address

To change the GPIB address of the instruments, change the following parameters in the "GPIB Setup" block.

Parameter

Description

Default

 

 

 

Hpib_sc

4155/4156 GPIB select code

7

 

 

 

Hpib_addr

4155/4156 GPIB address

17

 

 

 

Swm_sc

E5250A GPIB select code

7

 

 

 

Swm_addr

E5250A GPIB address

22

 

 

 

Agilent E5250A User’s Guide, Edition 9

9-35

Page 291
Image 291
Agilent Technologies Agilent E5250A manual Modifying the Program