Agilent E5250A User’s Guide, Edition 9 9- 23
Executing Sample Programs
HCI Measurement Program
Adding the Bias Source Control Routine
This sample program requires bias sources to apply stress to DUT. So, you must
modify the following bias source control subprograms to match y our bias s ources or
create new subprograms. The following subprogram is for bias sources that support
SCPI commands:
1. Get the HCI test program (SP_MUX_M) from the E5250A Program Disk by
using the computer:
GET "SP_MUX_M"
2. Delete the exclamation marks (!) from beginning of the following two program
lines. You can easily find these lines by using FIND statement.
! CALL Bias_source_on
: :
! CALL Bias_source_off
3. Delete the exclamation mark (!) from the beginning of the following lines in the
"Bias_source_on" and "Bias_source _off" subprograms.
Then modify the "OUTPUT" statements to match your Bias Sources. The
"SOUR XXX" SCPI command is used to control the output voltage of the Bias
Sources.
!Bias_source_on:SUB Bias_source_on
! !=================================== == == == == == == == == == == == ==
! ! Bias Source On
! !----------------------------------- -- -- -- -- -- -- -- -- -- -- -- --
! COM /Meas_info5/ REAL Vdstr,Vgstr,Vdd,Vbb, Gate_length,Gate_w
idth
! !----------------------------------- -- -- -- -- -- -- -- -- -- -- -- --
! OUTPUT 701;"SOUR "&VAL$(Vgstr) ! Set Gate Stress Voltage
! OUTPUT 702;"SOUR "&VAL$(Vdstr) ! Set Drain Stress Voltage
! OUTPUT 703;"SOUR 0.0" ! Set Substrate Stress Volta ge
!SUBEND
! !=================================== ========================
!Bias_source_off:SUB Bias_source_off
! !=================================== == == == == == == == == == == == ==
! ! Bias Source Off
! !----------------------------------- ------------------------
! COM /Meas_info5/ REAL Vdstr,Vgstr,Vdd,Vbb, Gate_length,Gate_w
idth
! !----------------------------------- ------------------------
! OUTPUT 701;"SOUR 0.0" ! Set Gate Voltage to 0.0V
! OUTPUT 702;"SOUR 0.0" ! Set Drain Voltage to 0.0V
! OUTPUT 703;"SOUR 0.0" ! Set Substrate Voltage to 0.0 V
!SUBEND
4. Save modified file to an empty formatted diskette as in following example. This
diskette is used for executing program. And the data measured and analyzed will
be saved to this diskette automatically during program execution.
RE-STORE "SP_MUX"