Executing Self-Test and Leak Test

Using the Self-Test Utility

Executing the Controller Test

To execute the controller test, press the CONTROLLER softkey.

If [F] result is displayed after the test, the E5250A is defective. Contact your nearest Agilent Technologies Service Center.

Executing the Front Panel Interface Test

To execute the front panel interface test, do the following:

 

1.

Press the FRONT PANEL softkey.

 

2.

Confirm that the following four LEDs on the E5250A front panel are blinking.

 

 

And press the Local/Self Test key on the E5250A front panel within 10 seconds.

 

 

Remote LED

 

 

System Fail LED

 

 

Fail LED

 

 

Local/Self Test LED

 

Fail Condition for Front Panel Interface Test

NOTE

If an LED does not blink after selecting the FRONT PANEL softkey, the LED is defective.

If you do not press the Local/Self Test key within 10 seconds, the Fail LED turns on.

If you pressed Local/Self Test key within 10 seconds, and the Fail LED still turns on, it means the Local/Self Test key is defective.

Contact your nearest Agilent Technologies Service Center.