Command Reference *STB?

*STB?

This query command reads the Status Byte Register (reads Master Summary Status bit, not Request for Service Message).

For bit6, this command reads MSS, not Request for Service (RQS). Refer to “Status Reporting Structure” on page 7-48.

Syntax

*STB?

 

 

 

 

 

Query response

register <newline><^END>

 

 

 

 

 

 

 

 

Parameter

 

 

Explanation

 

 

 

 

 

 

 

 

 

register

 

 

decimal integer value that is the sum of the binary-weighted values

 

 

 

 

 

for the set bits

Semantics

 

 

 

 

 

The following table shows the bits of the Status Byte Register, and the

 

binary-weighted decimal value of each bit.

 

 

 

 

 

 

 

bit

 

binary-

description

 

 

 

weight

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

1

not used. always 0.

 

 

 

 

 

 

 

 

 

1

 

 

2

not used. always 0.

 

 

 

 

 

 

 

 

 

2

 

 

4

not used. always 0.

 

 

 

 

 

 

 

 

 

3

 

 

8

not used. always 0.

 

 

 

 

 

 

 

 

 

4

 

 

16

MAV (Message Available summary-message)

 

 

 

 

 

 

 

 

 

5

 

 

32

ESB (Event Status Bit)

 

 

 

 

 

 

 

 

 

6

 

 

64

MSS (Master Summary Status)

 

 

 

 

 

 

 

 

 

7

 

 

128

not used. always 0.

 

 

 

 

 

 

Example

OUTPUT @Hp5250;"*STB?"

 

ENTER @Hp5250;A

Agilent E5250A User’s Guide, Edition 9

7-13

Page 181
Image 181
Agilent Technologies Agilent E5250A manual Stb?