Agilent E5250A User’s Guide, Edition 9 10- 9
Specifications
Supplemental Data when used with Agilent 4155/4156
Supplemental Data when used with Agilent 4155/4156
The below data shows degradation of Agilent 4155/4156 when used with Low
Leakage I-V Port.
Iout is output current in A, and Vout is output voltage in V.
Supplemental Data when used with Agilent 4155
Voltage Range Offset Accuracy (with MPSMU)
Current Range Offset Accuracy (with MPSMU)
Supplemental Data when used with Agilent 4156 (non-Kelvin connection)
Voltage Range Offset Accuracy (with HRSMU)
Current Range Offset Accuracy (with HRSMU)
Range Setting Measurement
2 V 960 μV + (0.5 × Iout) V 780 μV + (0.5 × Iout) V
Range Setting Measurement
1 nA 3 pA + (0.2 × Vout) pA 3 pA + (0.2 × Vout ) pA
10 nA 7 pA + (0.2 × Vout) pA 5 pA + (0.2 × Vou t) pA
100 nA 50 pA + (0.4 × Vout) pA 30 pA + (0.4 × Vout) pA
Range Setting Measurement
2 V 500 μV + (0.5 × Iout) V 280 μV + (0.5 × Iout) V
Range Setting Measurement
10 pA 0.4 pA + (0.2 × Vout) pA 0.1 pA + (0.2 × Vo ut ) pA
100 pA 0.4 pA + (0.2 × Vout) pA 0.1 pA + (0.2 × Vou t) p A
1 nA 0.7 pA + (0.2 × Vout) pA 0.5 pA + (0.2 × Vou t ) p A
10 nA 4 pA + (0.2 × Vout) pA 2 pA + (0.2 × Vou t) pA
100 nA 40 pA + (0.3 × Vout) pA 20 pA + (0.3 × Vout) pA