Specifications

Supplemental Data when used with Agilent 4155/4156

Supplemental Data when used with Agilent 4155/4156

The below data shows degradation of Agilent 4155/4156 when used with Low Leakage I-V Port.

Iout is output current in A, and Vout is output voltage in V.

Supplemental Data when used with Agilent 4155

Voltage Range Offset Accuracy (with MPSMU)

Range

Setting

Measurement

 

 

 

2 V

960 μV + (0.5 ⋅ Iout) V

780 μV + (0.5 ⋅ Iout) V

 

 

 

Current Range Offset Accuracy (with MPSMU)

Range

Setting

Measurement

 

 

 

1 nA

3 pA + (0.2 ⋅ Vout) pA

3 pA + (0.2 ⋅ Vout) pA

 

 

 

10 nA

7 pA + (0.2 ⋅ Vout) pA

5 pA + (0.2 ⋅ Vout) pA

 

 

 

100 nA

50 pA + (0.4 ⋅ Vout) pA

30 pA + (0.4 ⋅ Vout) pA

 

 

 

Supplemental Data when used with Agilent 4156 (non-Kelvin connection)

Voltage Range Offset Accuracy (with HRSMU)

Range

Setting

Measurement

 

 

 

2 V

500 μV + (0.5 ⋅ Iout) V

280 μV + (0.5 ⋅ Iout) V

 

 

 

Current Range Offset Accuracy (with HRSMU)

Range

Setting

Measurement

 

 

 

10 pA

0.4 pA + (0.2 ⋅ Vout) pA

0.1 pA + (0.2 ⋅ Vout) pA

 

 

 

100 pA

0.4 pA + (0.2 ⋅ Vout) pA

0.1 pA + (0.2 ⋅ Vout) pA

 

 

 

1 nA

0.7 pA + (0.2 ⋅ Vout) pA

0.5 pA + (0.2 ⋅ Vout) pA

 

 

 

10 nA

4 pA + (0.2 ⋅ Vout) pA

2 pA + (0.2 ⋅ Vout) pA

 

 

 

100 nA

40 pA + (0.3 ⋅ Vout) pA

20 pA + (0.3 ⋅ Vout) pA

 

 

 

Agilent E5250A User’s Guide, Edition 9

10-9

Page 303
Image 303