Programming the E5250A

Programming Examples

Program List:

10

! Example: Using Bias Mode

 

20

!

 

 

30

INTEGER Complete

 

40

DIM Channel$[100]

 

50

Channel$="(@10101:10108)"

 

60

!

 

 

70

! Put Instrument Initialization Routine here.

 

80

!

 

 

90

ASSIGN @Hp5250 TO 722

 

100 OUTPUT @Hp5250;"*RST"

 

110 OUTPUT @Hp5250;":ROUT:FUNC NCON"

 

120 OUTPUT @Hp5250;":ROUT:CONN:RULE ALL,FREE"

 

130 !OUTPUT @Hp5250;":ROUT:BIAS:PORT 1,7"

 

140 OUTPUT @Hp5250;":ROUT:BIAS:CHAN:ENAB:CARD 1"

 

150 OUTPUT @Hp5250;":ROUT:BIAS:STAT 1,ON"

 

 

 

 

 

 

 

 

Line

Description

 

 

Number

 

 

 

 

 

 

 

 

 

 

 

40

Declares data size of Channel$, which is used as channel_list

 

 

 

 

parameter.

 

 

 

 

 

 

 

 

50

Defines the desired channel_list.

 

 

 

 

 

 

 

 

70

Here, insert the initialization routine for your measurement

 

 

 

 

instrument.

 

 

 

 

 

 

 

 

100

Resets the E5250A.

 

 

 

 

 

 

 

 

110

Sets the E5250A to Normal channel configuration mode.

 

 

 

 

 

 

 

 

120

Sets the connection rule to FREE.

 

 

 

 

 

 

 

 

130

This line is not used (commented) for the E5255A. If you use the

 

 

 

 

E5252A instead of the E5255A, use this line, which defines AUX

 

 

 

 

INPUT HF1 to be the input Bias Port.

 

 

 

 

 

 

 

 

140

Bias enables all output ports of card 1 for the Bias Mode.

 

 

 

 

 

 

 

 

150

Turns on the Bias Mode. All output ports of card 1 are connected to

 

 

 

 

the input Bias Port.

 

 

 

 

 

Agilent E5250A User’s Guide, Edition 9

6-17

Page 153
Image 153
Agilent Technologies Agilent E5250A manual Example Using Bias Mode